Ecoflexmea24 – Multichannel Systems MEA Manual User Manual
Page 132
![background image](/manuals/338871/132/background.png)
Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany
Fon +49-7121-9 09 25- 0
Fax +49-7121-9 09 25-11
[email protected]
www.multichannelsystems.com
© 201
3 Multi Channel Systems MCS GmbH
Product information is subject to change
without notice.
EcoFlexMEA24
Flexible Microelectrode Array with 24 electrodes for use with 32-Channel Miniature Preamplifer
MPA32I for in vivo or in vitro applications.
Technical Specifications EcoFlexMEA24
Temperature compatibility
10 - 125 °C
Dimension (W x D)
36 mm x 25 mm
Thickness (region of electrodes)
50 μm
Base
material
Polyimide
(Kapton)
Contact
pads
and
leads
Gold
Electrode
diameter
80
μm
Interelectrode distance (centre to centre)
300 μm
Electrode
heights Planar
Electrode
type
Gold
Isolation
type
Polyimide
(Kapton)
Electrode impedance
Approximately 30 - 50 k
Electrode layout grid
10 x 2 + 4
Number of recording electrodes
23
Number of reference electrodes
2 internal reference electrodes
Number of ground electrodes
2 ground electrodes
MC_Rack
Version 4.1.1 and higher
Data Source Setup
Source Layout: Configuration, Total Number of Channels: 64,
Amplifier:
FA32I/S
or
FA64I/S,
MEA:
EcoFlexMEA24.
Channel map
Display: Layout tab > Default Map (EcoFlexMEA24.cmp).
Cleaning
Rinse with distilled water. EcoFlexMEAs are heat stable and can be autoclaved.
1......35 Pin Top GND
2.....36 Pin Bottom Electrodes
Top
35 33 31 29 27 25 23 21 19 17 15 13 11 9 7 5 3 1
36 34 32 30 28 26 24 22 20 18 16 14 12 10 8 6 4 2
Bottom
Top
Connect the EcoFlexMEA24 directly to a 32-Channel Miniature Preamplifier.
The electrode field is at the bottom of the EcoFlexMEA24.
Insert the EcoFlexMEA24 into the MPA32I with the electrode field up.