Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany
Fon +49-7121-9 09 25- 0
Fax +49-7121-9 09 25-11
[email protected]
www.multichannelsystems.com
© 201
3 Multi Channel Systems MCS GmbH
Product information is subject to change
without notice.
FlexMEA36
Technical Specifications FlexMEA36
Temperature compatibility
10 - 40 °C
Dimension (W x D)
31 mm x 18.5 mm
Thickness of the electrode field
12 μm
Weight
<
1
g
Base
material
Polyimide
(2611)
foil
Contact pads and track material
Gold
Electrode
diameter
30
μm
Interelectrode distance (centre to centre)
300 μm
Perforation
Diameters
of
the
holes
30
μm
Electrode
height Planar
Electrode
type
TiN
electrodes
(Titanium
nitride)
Isolation
type
Polyimide
(2610)
foil
Electrode
impedance
Approximately
50
k
Electrode
layout
grid
6
x
6
Number of recording electrodes
32
Number of reference electrodes
2 internal reference electrodes
Number
of
ground
electrodes
2
MC_Rack
Source layout in “Data Source Setup”
Configuration, 64 channels, Amplifier: FA64I/S or FA32I/S,
MEA:
FlexMEA36
(FlexMEA36.cmp).
Channel Map
“Default Map” in Layout tab of Data Display.
Cleaning
Rinse with distilled water, optional with ethanol 70%.
Do not autoclave or sterilize FlexMEAs by heat. These MEA types are not heat-stable
and will be irreversibly damaged! Please do not use ultrasonic bath for cleaning.
Flexible Microelectrode Array with 36 electrodes for use with 32-Channel Miniature
Preamplifier MPA32I-Flex or with the ADPT-FM-32 adapter and the standard MPA32I.
28
22
11
5
29
27
21
12
6
4
30
23
20
13
10
3
31
24
17
16
9
2
32
25
18
15
8
1
26
19
14
7
28
22
11
5
29
27
21
12
6
4
30
23
20
13
10
3
31
24
17
16
9
2
32
25
18
15
8
1
26
19
14
7
2200
31000
5000
3000
350
500
18500
Dimensions in μm