Technical Specifications 60pMEA100/30iR-Ti
Temperature compatibility
10 - 50 °C
Dimension (W x D x H)
49 mm x 49 mm x 1 mm
Base
material
Polyimide
foil
(2611)
on
glass
or
ceramic
carrier
Perforation:
Diameter of innermost area
2 mm
Total area of holes
19% (according to 2 mm)
Diameter of holes
5, 6, 7.5, 9, 10, 12.5, 15, 17.7, 20
Contact
pads
TiAu
(Titan,
Gold)
Track material
TiAuTi (Titan, Gold, Titan)
Electrode
diameter 30
μm
Interelectrode distance (centre to centre)
100 μm
Electrode
height
Planar
Electrode
type
TiN
(Titanium
nitride)
Isolation
type
Polyimide
foil
(2610)
isolator
Electrode impedance
30 - 50 k
Electrode
layout
grid
6
x
10
Number
of
electrodes
60
Reference
electrodes
with
internal
reference
(iR)
MC_Rack
Source layout in “Data Source Setup”
2 dim. (MEA)
Channel
map
pMEA
6x10.cmp
Cleaning
Rinse with distilled water. Do not use ultrasonic bath!
These pMEAs are not heat stable, and should not be autoclaved!
MEA perfusion chamber
(w/o)
Without ring
(gr)
Glass ring: ID +/- 19 mm, OD 24 mm, height 6 / 12 mm
(pr)
Plastic ring without thread: ID 26.5 mm, OD 30 mm, height 6 / 3 mm
(pr-T)
Plastic ring with thread: ID 26 mm, OD 30 mm, height 6 / 15 mm
Perforated Microelectrode Array with 6 x 10 electrode layout grid. Perforated MEAs for use with MEA1060 amplifiers
equipped with a perfusion ground plate (PGP) or a MEA2100 headstage with a perfusion element (PE). The perforation
allows a perfusion of the tissue from both sides of the pMEA.
60pMEA100/30iR-Ti
60pMEA100/30iR-Ti