beautypg.com

Ecoflexmea36 – Multichannel Systems MEA Manual User Manual

Page 135

background image

Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany

Fon +49-7121-9 09 25-

0

Fax +49-7121-9 09 25-11

[email protected]
www.multichannelsystems.com

© 201

3 Multi Channel Systems MCS GmbH

Product information is subject to change without
notice.

EcoFlexMEA36

Electrode layout

GND 2

GND 1

29

31

1

3

27

28

30

2

4

5

25

26

32

8

6

7

23

22

24

16

10

9

21

20

18

14

12

11

19

17

15

13

REF

REF

50

m

GND 1 is a large ground electrode connected to pin 1 of the MPA32I input connector. GND 2 is a second

ground electrode connected to pin 36. The REF electrodes are reference electrodes connected to pin 2

and 35, respectively. Both ground inputs and both reference electrode inputs are equal, that is, they are

connected to each other inside the standard MPA32I. Please see the MPA32I manual for details.

The numbers in the electrodes are the recording channel numbers that refer to the channel numbers

in the MC_Rack program. Please make sure that you have selected “Configuration” in the “Channel Layout”

under “Data Source Setup” with a total number of 64 channels. Deselect the check box “Digital input channel”

if you do not need it, otherwise it could be that one recording channel is missing! In “Amplifier”, please choose

FA32I/S or FA64I/S and in “MEA” EcoFlexMEA36. In Layout tab of the display, please click ”Default Map”.

The letter digit code below is the electrode identifier and refers to the position of the electrode in the grid.

Important: The MC_Rack channel map is constructed by looking on the back side of the electrodes,

because the FlexMEA electrodes are placed on the preparation upside down!

Warning: The device may only be used together with the MPA32I from Multi Channel Systems

MCS GmbH, and only for the specified purpose. Damage of the device and even injuries can

result from improper use.

!

300

m

A2 A3 A4 A5

B1 B2 B3 B4 B5 B6

C1 C2 C3 C4 C5 C6

D1 D2 D3 D4 D5 D6

E1 E2 E3 E4 E5 E6

F2 F3 F4 F5

300

m

If you use a MEA64-System with more than one MPA32I, the signal collector SC2x32 leads the output

channels of the second amplifier to channel numbers 33 to 64. Please read the SC2x32 data sheet.