Technical Specifications 60-4QMEA1000
Temperature compatibility
0 - 125 °C
Dimension (W x D x H)
49 mm x 49 mm x 1 mm
Base
material
Glass
Track
material
Ti
(Titanium)
Contact
pads
TiN
(Titanium
nitride)
Electrode
diameter
30
μm
Interelectrode distance (centre to centre)
200 μm inside of the quadrants
1000
μm between the quadrants
500
μm from quadrants to the center line
Electrode
height Planar
Electrode
type
TiN
(Titanium
nitride)
Isolation
type
Silicon
nitride
500
nm
(PEVCD)
Electrode impedance
Approximately 30 - 50 k
Electrode layout grid
4 x (1 x 4 + 1 x 5 + 1 x 4) + center line 1 x 7
Number of recording electrodes
59
Number of reference electrodes
1 internal reference electrode (iR)
MC_Rack:
Source layout in “Data Source Setup”
2 dim (MEA)
Channel
map
4QMEA.cmp
MEA perfusion chamber
(w/o) Without ring
(gr)
Glass ring: ID +/- 19 mm, OD 24 mm, height 6 / 12 mm
(pr)
Plastic ring without thread: ID 26.5 mm, OD 30 mm, height 6 / 3 mm
(pr-T) Plastic ring with thread: ID 26 mm, OD 30 mm, height 6 / 15 mm
60-4QMEA1000
60-4QMEA1000iR-Ti
Four Quadrants Microelectrode Array
200
μ
m 200
μ
m 1000
μ
m
500
μ
m 500
μ
m
200
μ
m 200
μ
m 200
μ
m 200
μ
m
1000
μ
m
1000 μm
30 μm
500 μm