Technical Specifications 60-4QMEA1000

Temperature compatibility

0 - 125 °C

Dimension (W x D x H)

49 mm x 49 mm x 1 mm

Base

material

Glass

Track

material

Ti

(Titanium)

Contact

pads

TiN

(Titanium

nitride)

Electrode

diameter

30

μm

Interelectrode distance (centre to centre)

200 μm inside of the quadrants

1000

μm between the quadrants

500

μm from quadrants to the center line

Electrode

height Planar

Electrode

type

TiN

(Titanium

nitride)

Isolation

type

Silicon

nitride

500

nm

(PEVCD)

Electrode impedance

Approximately 30 - 50 k

Electrode layout grid

4 x (1 x 4 + 1 x 5 + 1 x 4) + center line 1 x 7

Number of recording electrodes

59

Number of reference electrodes

1 internal reference electrode (iR)

MC_Rack:
Source layout in “Data Source Setup”

2 dim (MEA)

Channel

map

4QMEA.cmp

MEA perfusion chamber

(w/o) Without ring

(gr)

Glass ring: ID +/- 19 mm, OD 24 mm, height 6 / 12 mm

(pr)

Plastic ring without thread: ID 26.5 mm, OD 30 mm, height 6 / 3 mm

(pr-T) Plastic ring with thread: ID 26 mm, OD 30 mm, height 6 / 15 mm

60-4QMEA1000

60-4QMEA1000iR-Ti

Four Quadrants Microelectrode Array

200

μ

m 200

μ

m 1000

μ

m

500

μ

m 500

μ

m

200

μ

m 200

μ

m 200

μ

m 200

μ

m

1000

μ

m

1000 μm

30 μm

500 μm

200 μm