Agilent Technologies E1300B User Manual
Page 223
![background image](/manuals/33541/223/background.png)
Error Messages and Causes
Code
Message
Cause
-101
Invalid character
U nrecognized character in specified parameter.
- 102
Syntax error
Command is missing a space or comma between
parameters
- 103
Invalid separator
Command parameter is separated by some
character other than a comma.
- 104
D ata type error
The wrong data type (i.e. number, character,
string expression) was used when specifying a
parameter.
- 108
Parameter not allowed
Parameter specified in a command which does
not require one.
- 109
Missing parameter
No parameter specified in the command in
which a parameter is required.
- 113
U ndefined header
Command header was incorrectly specified.
- 123
Numeric overflow
A parameter specifies a value greater than the
command allows.
- 128
Numeric data not allowed
A number was specified for a parameter when a
letter is required.
- 131
Invalid suffix
Parameter suffix incorrectly specified
(e.g. .5SECO ND rather than .5S or .5SEC).
- 138
Suffix not allowed
Parameter suffix is specified when one is not
allowed.
- 141
Invalid character data
The discrete parameter specified is not allowed
(e.g. TRIG :SO U R INT - INT is not a choice.)
- 178
Expression data not allowed
A parameter other than the channel list is
enclosed in parentheses.
- 211
Trigger ignored
Trigger occurred while the Pacer is in the idle
state, or a trigger occurred from a source other
than the specified source.
- 222
D ata out of range
The parameter value specified is too large or too
small.
- 224
Illegal parameter value
The numeric value specified is not allowed.
- 240
H ardware error
H ardware error detected during power-on cycle.
Return multimeter to Agilent for repair.
- 310
System error
If caused by *D MC, then macro memory is full.
- 350
Too many errors
The error queue is full as more than 30 errors
have occured.
- 410
Q uery interrupted
D ata is not read from the output buffer before
another command is executed.
- 420
Q uery unterminated
Command which generates data not able to
finish executing due to a multimeter
configuration erorr.
- 430
Q uery deadlocked
Command execution cannot continue since the
mainframe’s command input, and data output
buffers are full. Clearing the instrument restores
control.
1500
External trigger source
already allocated
"Event In" signal already allocated to another
instrument such as a Switchbox.
2002
Invalid logical address
A value less than 0 or greater than 255 was
specified for logical address.
2003
Invalid word address
An odd address was specified for a 16 bit read or
write. Always use even addresses for 16 bit
(word) accesses.
2005
No card at logical address
A non-existent logical address was specified with
the VXI:READ ? or VXI:WRITE command.
2101
Failed D evice
VXI device failed its self test.
2102
U nable to combine device
D evice type can not be combined into an
instrument such as a scanning voltmeter or a
switchbox.
2103
Config warning, D evice
driver not found
ID of device does not match list of drivers
available. Warning only.
2105
Config error 5, A24 memory
overflow
More A24 memory installed in the mainframe
than can be configured into the available A24
memory space.
2108
Config error 8, Inaccessible
A24 memory
A24 memory device overlaps memory space
reserved by the mainframe’s operating system.
Table B-2. Error Messages and Causes
Error Messages B-3