Agilent Technologies E1300B User Manual
Page 135
:BOOT[:WARM]
DIAGnostic:BOOT[ :WARM] causes the System Instrument to restart (re-boot)
using the current configuration stored in non-volatile memory. The effect is the
same as cycling power.
Comments
•
The System Instrument goes through its power-up self tests.
•
The non-volatile system state is used for configuration wherever
applicable.
•
Related Commands: DIAG:BOOT:COLD
Example
Booting the System Instrument (warm)
DIAG:BOOT:WARM
force boot
:COMMunicate
:SERial[0][:OWNer]
DIAGnostic:COMMunicate:SERial[ 0] [ :OWNer] < owner > Allocates the
built-in serial interface to the System Instrument, the optional IBASIC
interpreter, or to neither.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
owner
discrete
SYSTem| IBASic| NONE
none
Comments
•
While the serial interface is allocated to the Command Module
(SYSTem), it can function as the mainframe user interface when
connected to a terminal or computer running terminal emulation
software.
•
When the built-in serial interface is allocated to IBASIC, it is controlled
only by IBASIC. The serial interface is given a select code of 9, and any
RS-232 device connected to the (Command Module) RS-232 port is
programmed accordingly.
•
If the built-in serial interface is not needed, specifying NONE will release
memory for use by other instruments.
•
Once the new serial interface owner has been specified
(DIAG:COMM:SER:OWN), the change will not take effect until you
re-boot (warm) the system.
•
Related Commands: DIAGnostic:COMMunicate:SERial[:OWNer]
Example
Give the serial interface to IBASIC.
DIAG:COMM:SER IBAS
Note; :OWNer is implied
DIAG:BOOT:WARM
Complete the allocation
DIAGnostic :BOOT[:WARM]
System Instrument Command Reference 7-7