Agilent Technologies E1300B User Manual
Page 202
![background image](/manuals/33541/202/background.png)
Remote
Sets the GPIB remote enable line (REN) true which places an instrument in the
remote state.
Comments
•
The REMOTE 709ss (ss = secondary address) command places the
instrument in the remote state. The REMOTE 7 command, does not, by
itself, place the instrument in the remote state. After sending the
REMOTE 7 command, the instrument will only go into the remote state
when it receives its listen address.
•
In most cases, you will only need the REMOTE command after using the
LOCAL command. REMOTE is independent of any other GPIB activity
and toggles a single bus line called REN. Most controllers set the REN
line true when power is applied or when reset.
Examples
REMOTE 7
Sets GPIB REN line true
REMOTE 70900
Sets REN line true and
addresses System Instrument
Serial Poll (SPOLL)
The SPOLL command, like the *STB? Common Command, returns the
weighted sum of all set bits in an instrument’s Status Register (status byte).
Refer to "The Status Register" earlier in this chapter for a table showing the
contents of the Status Register.
Comments
•
The SPOLL command differs from the *STB? command in that SPOLL
clears bit 6 (RQS). Executing *STB? does not clear bit 6.
Examples
10 P= SPOLL (70900)
Sends Serial Poll, places
response into P
20 DISP P
Displays response
30 END
7-74 System Instrument Command Reference