[:IMMediate]
TRIGger:IMMediate will cause a trigger cycle to occur immediately, provided
that the trigger system has been initiated (INITiate).
Comments
•
Related Commands: ABORt, INITiate
•
*RST Condition: This command is an event and has no *RST condition.
Example
Triggering the Pacer.
TRIG:SOUR HOLD
trigger source is TRIG
command
SOUR:PULS:COUN 1E3
output 1000 Pacer pulses
SOUR:PULS:PER .1 S
pulse period set to .1 second
TRIG:DELAY .75 S
start Pacer .75 sec after trigger
INIT
go to Wait For Trigger state
TRIG
trigger Pacer to output pulses.
:SLOPe
TRIGger:SLOPe < slope> is for SCPI compatibility. The mainframe’s "Event
In" signal only triggers on a negative going edge.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
slope
discrete
NEGative
none
Comments
•
Trying to set
…
SLOPe to other than NEG will generate an error.
•
Related Commands: ABORt, INITiate,
:SLOPe?
TRIGger:SLOPe? returns the current trigger slope setting. Since the
mainframe’s "Event In" signal only triggers on a negative going edge,
TRIG:SLOP? will always return "NEG".
:SOURce
TRIGger:SOURce < trig_source> configures the trigger system to respond to
the specified source.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
trig_source
character
BUS| EXT| HOLD| IMM
none
TRIGger[:IMMediate]