Agilent Technologies E1300B User Manual
Page 178
:TIME
SYSTem:TIME < hour> ,< minute> ,< second> sets the E1300/E1301
mainframe’s internal clock.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
hour
numeric
must round to 0 to 23
none
minute
numeric
must round to 0 to 59
none
second
numeric
must round to 0 to 60
none
Comments
•
Related Commands: SYST:DATE, SYST:DATE?, SYST:TIME?
•
*RST Condition: *RST does not change the Command Module’s real
time clock.
Example
Setting the system time
SYST:TIME 14,30,20
set 2:30:20 PM
:TIME?
SYSTem:TIME? [ MAX| MIN,MAX| MIN,MAX| MIN] returns:
•
When no parameter is sent; the current system time in the form
+ HH,+ MM,+ SS, where HH can be 0 through 23 hours, MM can be 0
through 59 minutes, and SS can be 0 through 60 seconds.
•
When parameters are sent; the minimum or maximum allowable values
for each of the three parameters. The parameter count must be three.
Example
Querying the system time
SYST:TIME?
ask for current time
input values of hour,min,sec
read back time
:VERSion?
SYSTem:VERSion? Returns the SCPI version for which this instrument
complies.
Comments
•
The returned information is in the format: YYYY.R; where YYYY is the
year, and R is the revision number within that year.
•
Related Commands: *IDN?
Example
Determine compliance version for this instrument.
SYST:VERS?
enter statement
Statement enters 1990.0
SYSTem:TIME
7-50 System Instrument Command Reference