Agilent Technologies E1300B User Manual
Page 187
registers) which are present on the device. -1 indicates that the device has
no A16 memory.
•
A24 memory offset: an integer between -1 and 16777215 inclusive.
Indicates the base address for any A24 registers which are present on the
device. -1 indicates that the device has no A24 memory.
•
A32 memory offset: an integer between -1 and 4294967295 inclusive.
Indicates the base address for any A32 registers which are present on the
device. -1 indicates that the device has no A32 memory.
•
A16 memory size: an integer between -1 and 65535 inclusive. Indicates
the the number of bytes reserved for any A16 registers (other than the
VXIbus defined registers) which are present on the device. -1 indicates
that the device has no A16 memory.
•
A24 memory size: an integer between -1 and 16777215 inclusive. Indicates
the number of bytes reserved for any A24 registers which are present on
the device. -1 indicates that the device has no A24 memory.
•
A32 memory seze: an integer between -1 and 4294967295 inclusive.
Indicates the number of bytes reserved for any A32 registers which are
present on the device. -1 indicates that the device has no A32 memory.
•
Slot number: an integer between -1 and the number of slots which exist in
the cage. -1 indicates that the slot which contains this device is unknown.
•
Slot 0 logical address: an integer between -1 and 255 inclusive. -1
indicates that the Slot 0 device associated with this device is unknown.
•
Subclass: an integer representing the contents of the subclass register. -1
indicates that the subclass register is not defined for this device.
•
Attribute: an integer representing the contents of the attribute register. -1
indicates that the attribute register is not defined for this device.
•
Manufacturer’s Specific Comments. Up to 80 character string contains
manufacturer specific data in string response data format. This field is
sent with a 488.2 string response data format, and will contain the
instrument name and its IEEE 488.1 secondary address unless a start-up
error is detected. In that case, this field will contain one or more error
codes in the form "CNFG ERROR: n, m, ...,z" . See Appendix B, Table
B-3 for a complete list of these codes.
Comments
•
Related Commands: VXI:SEL, VXI:CONF:INF:ALL?,
VXI:CONF:LADD?
Example
Query information on logical address 0.
VXI:SEL 0
select the logical address
VXI:CONF:INF?
ask for data
enter statement
return data
VXI :CONFigure :INFormation?
System Instrument Command Reference 7-59