beautypg.com

GF Signet 2759 pH-ORP System Tester User Manual

Signet 2759 ph/orp simulator and system tester, Mv value, English

background image

Features:

A) Power OFF Button

B) Output simulation buttons and indicators.

Simulate pH and ORP output at fi ve fi xed values:
pH 4, pH 7, pH 10, –700 mV and +700 mV.

Pressing one of these buttons turns the 2759 on.

C) Low battery indicator

D) High Ω switch:

• Adds 1000 MΩ resistance in series with output.

• Simulates high impedance of pH electrodes.
Used to verify proper preamplifi er operation.

E) Adapter cable:

PN 3-2759.391 for use with the 2750 or 2760.

F) Bypass adapter cable:

PN 3-2759.390 (included with 2759) to connect directly to an instrument.

G) Mode selector switch:

Trigger automatic sensor recognition software in Signet pH/ORP instrumentation.

• Top = 1K for a Signet 8900 or 9900 instrument needing PT1000 temperature compensation input.

• Middle = 10K for ORP simulation.

• Bottom = 3K for Signet 5700 and 8750 instruments needing a 3K temperature compensation input.

H) 2750 Sensor Electronics

I) 2760

Preamplifi er

Signet 2759 pH/ORP Simulator and System Tester

1. Description

The Signet 2759 pH/ORP Simulator is a battery-powered millivolt generator that simulates pH values of 4, 7 and 10, plus ORP values of
± 700 mV. This device is useful as a troubleshooting aid and for general verifi cation of system operation. It is not a substitute for periodic
system calibration with pH buffers or test solutions. Accessory adapter cables (sold separately) enable the 2759 to connect directly to
2750 pH/ORP Sensor Electronics or 2760 pH/ORP Preamplifi er. The adapters include a selector switch for pH or ORP simulation. The
switch triggers automatic sensor-recognition software in Signet pH/ORP instrumentation.

2. Specifi cations

mV output accuracy: ............. ± 0.6 mV (± 0.01 pH)

pH system temperature simulation:
w/2750 adapter: ............... 1.1 KΩ = 25 ºC (± 4 ºC)
w/2760 adapter: ............... 3 KΩ

= 25 ºC (± 4 ºC)

High Ω resistor value: ........... 1000 MΩ

Battery: ................................. 9V alkaline

Life: 400 hours

Dimensions: .......................... 100 x 75 x 23 mm

(3.94 x 2.95 x 0.91 in.)

Weight: ................................. 120 grams (5 oz.)

+177

mV

0

mV

-177

mV

-700

mV

+700

mV

mV Value

1. Description
2. Specifi cations
3. Troubleshooting ProcessPro, ProPoint and 2760 Preamplifi er
4. Troubleshooting 2750 pH/ORP Electronics
5. pH and ORP System Response Chart

3-2759.090 Rev. F 10/12 English

*3-2759.090*

English

+GF+

A

B

C

D

E

F

G

I

H

10K

ORP

1K

3K

V -

4-20

mA

T -

T +

V +

Iso. Gnd

mV Input

Signet

Instrument

China RoHS
For more information go to www.gfsignet.com