4. Technical Data Signet 2650 DryLoc Amperometric Electronics
Electrical
Input Specifi cations:
Sensor: Raw signal
Temperature: PT1000 RTD
Output Specifi cations:
Digital (S
3
L):
Serial ASCII, TTL level 9600 bps
Maximum Cable Length:
30 m (100 ft)
Power Supply Input:
Digital (S
3
L) mode:
5 to 6.5 V ± 10% , 3 mA max
Environmental
Storage Temperature:
–20 °C to 85 °C (–4 °F to 185 °F)
Operating Temperature:
0 °C to 85 °C (32 °F to 185 °F). Electronics only
Relative Humidity:
0 to 95%, non-condensing (no electrode connected)
Enclosure Requirements:
NEMA 4X/IP65 with electrode connected
Standards and Approvals
CE
Manufactured under ISO 9001 and ISO 14001
China RoHS (Go to www.gfsignet.com for details)
General
Compatible Sensors:
All Signet Amperometric DryLoc Sensors
Compatible Instruments:
Signet 3-8630-3P Chlorine Transmitter
Mounting:
DryLoc connection
Materials:
Valox
®
(PBT)
Cable:
4.6 m (15 ft) 3 conductor shielded, 22 AWG
Shipping Weight:
0.64 kg (1.14 lb)
Performance
Electronics Accuracy:
< 5 nA or 1% of reading, whichever is greater
@ 25 °C over full input range
Temperature:
± 1.0 °C (PT1000) over full operation range
(when calibrated at ambient temperature)
System Response:
500 ms (update rate)
Operational Range:
± 450 mV
Resolution:
0.1 nA
* Valox
®
is a registered trademark of SABIC Innovative Plastics
3. Wiring
Chlorine Digital (S
3
L) Wiring
•
Remove approximately 10 mm (0.4 in.) of insulation and tin each conductor before inserting into terminals.
•
Refer to the wiring diagram above to connect the 2650 cable to the terminals on the
8630
Chlorine
Transmitter.
•
Maximum cable length is 30 m (100 ft).
•
Cable should have noise immunity for the sensor signal of up to 30 m (100 ft).
•
Cable capacitance must be less than 30 pF/foot.
Gnd
I/O
V
+
15
14
13
Signal Ground
Digital (S3L) data
+5 VDC
White
Red
Black
Shield
35.0 mm
(1.38 in.)
61.0 mm
(2.40 in.)
101 mm
(4.0 in.)