13.12 SELF-TEST function class

The SELF-TEST function class is comprised of the functions relating to the self-test of the sensor.
The diagnostic functions of the transmitter, which monitor the proper functioning of the electronics and
the software, are always active and cannot be switched off. The excitation current can be monitored in
addition.

SELF-TEST

SELF-TEST

Sensor test

on / off

Sensor test

on / off

Max. deviation of

excitation

Max. deviation of

excitation

SELF-TEST

functional class

Sensor test

[off]

_______________

off
on

Max. deviation

020 %

Self-test

calibration

Self-test

calibration

Calibration

[no]

_______________

no

yes

Monitoring

sensor

Monitoring

sensor

S1 7749 S2 7812

E 1280


The excitation current of each sensor in the system individually depends on the sensor itself, the fluid and
the installation conditions. If the excitation currents changes while the fluid remains the same, conclu-
sions may be drawn for e.g. potential wear and tear, viscosity changes or air bubbles. The operator has
the possibility of defining a “normal state” (“Self-test calibration”) and setting the limit for a permissible
deviation. This function is deactivated in the device when delivered.

CMB & CT OPERATING MANUAL
Part No 06EN003384 rev. 1.6

Page 89 of 106