13.7 TEMPERATURE functional class

The TEMPERATURE functional class is comprised of the functions that affect the lower- and the upper-
range value and the processing of the measured temperature. The additional service functions will not be
described in these instructions. Modifications can only be made in Programming mode (see 12.3
Operating modes), which means that a correct
password (see 12.4.3.3 Passwords, 13.2 PASSWORD
functional class) must be entered.

TEMPERATURE

TEMPERATURE

Temperature

lower range val.

Temperature

lower range val.

Temperature

upper range val.

Temperature

upper range val.

TEMPERATURE

functional class

Temperature

0% = - 010 °C

Temperature

100% = + 100 °C

Temperature limit

MIN

Temperature limit

MIN

MIN. temperature

- 050 °C

Temperature limit

MAX

Temperature limit

MAX

MAX. temperature

+250 °C

Max. measured

temperature

Max. measured

temperature

Temperature

unit

Temperature

unit

Temperature in

[°C]

_____________

°C

°F

K

permissible +250 °C

measured +197°C


Page 72 of 106

CMB & CT OPERATING MANUAL