Fluid Components International CMB User Manual
Page 101
Display (English)
Display (German)
Description
Possible cause of error and
remedy
MAX
for temperature.
and the limit values if necessary.
Temperature < MIN Temperatur The measured temperature is Product temperature is too low; params inconsist Parameter inkons. Parameter is inconsistent. Check the parameter settings. The set parameters are contra- Example: Upper-range value, missing EEPROM ext EEPROM fehlt The data memory module (DSB) Insert the data storage module wrong EEPROM falsches EEPROM EEPROM of a former model (e.g. internal communi- interne Kommunika- Communication between control Contact the device ven- Information: Error message: “Parameter is inconsistent” (system error 0x0400)? To generate a list of the inconsistencies, first enter a valid password and then an inva- CMB & CT OPERATING MANUAL Page 101 of 106
below the set lower-range value
for temperature.
adjust the temperature range
and the limit values if necessary.
dictory.
pulse value and pulse duration
must be matched in such a way
that the combination fits for all
measured values.
with the calibration data of the
sensor and the customer-specific
settings of the transmitter is not
plugged-in.
(DSB/UMF33) in the correspond-
ing receptacle on the CPU
printed board UMC-30.
UMC2 or UMF) has been
plugged-in as memory module.
cation faulty
tion gestört
unit and transmitter is faulty.
dor/customer service depart-
ment.
lid password. The control unit will show a list of current errors (only once). The opera-
tor can then correct the inconsistent settings after entering a valid password.
Part No 06EN003384 rev. 1.6