Display of system error – Fluid Components International CMB User Manual
Page 102
15.3.2 Display of system error
System errors consist of the message text “system error” and a 5-digit number in hexadecimal code. The
meaning of the individual error codes is described in the following table. If several errors occur at the
same time, the hexadecimal sum of the individual errors will be displayed. The errors are coded in such a
way that the individual errors can be easily identified. The sums are unique.
Descriptor label
(never displayed)
Descriptor label
(never displayed)
Constant/
display
Description
System failure Div0
SystemfehlerDiv0
0x00001 Arithmetical error/division by zero
System failure
IntEEProm
SystemfehlerIntEEProm 0x00002
Transmitter data checksum is faulty;
reinitialization is necessary.
SystemfehlerPruefsumme
0x00004
Sensor data checksum is faulty.
SystemfehlerLeeresEEPROM
0x00008
Ext. EEPROM is present but empty
(no content).
SystemfehlerEEPROM
0x00010 Value could not be stored/read out.
System failure Phase SystemfehlerPhase
0x00020 Phase measurement/mass flow is
faulty.
System failure fre-
quency
SystemfehlerFrequenz 0x00040
Frequency measurement/density
measurement is faulty.
System failure DSP
Version
Systemfehler DSP Version
0x00080 DSP firmware is outdated (not ad-
justed to the transmitter operating
system)
SystemfehlerZeitkonstante
0x00100
Initialization of time constants failed.
SystemfehlerMesswert
0x00200
Faulty calculation of measured value
System failure Pa-
rameter
SystemfehlerParameter
0x00400 Settings are inconsistent.
SystemfehlerRAMPrüfsumme
0x00800
Defective main memory, inconsistent
checksum (custody transfer opera-
tion)
SystemfehlerFlashPrüfsumme
0x01000
Defective program memory, inconsis-
tent checksum
Page 102 of 106
CMB & CT OPERATING MANUAL