beautypg.com

Display of system error – Fluid Components International CMB User Manual

Page 102

background image

15.3.2 Display of system error

System errors consist of the message text “system error” and a 5-digit number in hexadecimal code. The
meaning of the individual error codes is described in the following table. If several errors occur at the
same time, the hexadecimal sum of the individual errors will be displayed. The errors are coded in such a
way that the individual errors can be easily identified. The sums are unique.

Descriptor label
(never displayed)

Descriptor label
(never displayed)

Constant/

display

Description

System failure Div0

SystemfehlerDiv0

0x00001 Arithmetical error/division by zero

System failure
IntEEProm

SystemfehlerIntEEProm 0x00002

Transmitter data checksum is faulty;
reinitialization is necessary.

SystemfehlerPruefsumme

0x00004

Sensor data checksum is faulty.

SystemfehlerLeeresEEPROM

0x00008

Ext. EEPROM is present but empty
(no content).

SystemfehlerEEPROM

0x00010 Value could not be stored/read out.

System failure Phase SystemfehlerPhase

0x00020 Phase measurement/mass flow is

faulty.

System failure fre-
quency

SystemfehlerFrequenz 0x00040

Frequency measurement/density
measurement is faulty.

System failure DSP
Version

Systemfehler DSP Version

0x00080 DSP firmware is outdated (not ad-

justed to the transmitter operating
system)

SystemfehlerZeitkonstante

0x00100

Initialization of time constants failed.

SystemfehlerMesswert

0x00200

Faulty calculation of measured value

System failure Pa-
rameter

SystemfehlerParameter

0x00400 Settings are inconsistent.

SystemfehlerRAMPrüfsumme

0x00800

Defective main memory, inconsistent
checksum (custody transfer opera-
tion)

SystemfehlerFlashPrüfsumme

0x01000

Defective program memory, inconsis-
tent checksum


Page 102 of 106

CMB & CT OPERATING MANUAL