6.0 MODELS WITH DEVICENET

6.1 DEVICENET POWER UP SEQUENCE

Applying power to a DeviceNet CMX initiates a power-up sequence that
initializes the embedded Vacuum Measurement system along with performing a
Module and Network test. The real time status of these tests is indicated on the
Module and Network LEDs on the top of the device.

After power is applied, the CMX flashes the Module and Network LEDs
according to the following sequence. If this sequence does not occur, a problem
exists with module power or with the device itself.

Module LED Power-up Sequence: Off, Green (250 mSec), Red (250 mSec).

Network LED Power-up Sequence: Off, Green (250 mSec), Red (250 mSec).

After the LED flash sequence has completed, the Module and Network LEDs will
display the real time status of the Module and Network respectively. The
following tables describe the status LED behavior:

Table 11: DeviceNet: MODULE LED

Table 12: DeviceNet: NETWORK LED

6.2 DEVICENET AND NETWORKING ISSUES AND SOLUTIONS

There are a number of initialization failure scenarios that can occur with
DeviceNet. These issues include wrong default assembles or produced and
consumed paths, incorrect address or baud rate settings, network wiring,
network termination, and network power supply problems. Several of the more
common initialization problems are:

LED STATE

PROBLEM

SOLUTION

Off

Power off

No power applied to the device.

Flashing Green-
Red

Device self-test

Device is in self-test. The DeviceNet
specifications defines blink rates and
patterns for the Modules Status LED
during self-test.

Green

Unit is operating normally

No corrective action required.

Red

Unrecoverable fault

Device has detected an unrecoverable
fault.

LED STATE

PROBLEM

SOLUTION

Off

Power off

No power applied to the device.

Flashing Green-
Red

Device self-test

Device is in self-test.

Flashing Green

On-line, not connected

Device is operating normal. It is on-line but no connection has
been established.

Green

On-line, connected

Device is operating normal. It is on-line and one or more
connections are established.

Flashing Red

Connection time out

One or more connections have timed out.

Red

Unrecoverable fault

Device cannot communicate on the network.
(Duplicate Macld or Bus-off condition)