beautypg.com

0 _description, Initial warm-up - analog (typical), Steady state - analog (typical) – Brooks Instrument XacTorr CMX160 User Manual

Page 16: Initial warm-up - devicenet, Typical), Steady state - devicenet (typical), Devicenet) w/9-pin d-sub female (analog)

background image

p.10

2.9 PRODUCT SPECIFICATIONS

Table 3: Digital CM Specifications

1

Includes hysteresis, linearity, and repeatability within the calibrated range at 21°C.

2

Ambient Operating Temperature with 50 linear feet of airflow = 15-45°C.

3

15 Pin Male D-Sub Model requires an additional 60mA to power the internal relays (if energized).

4

Process relays are UL

®

recognized components rated for 100,000 cycles at 1A, 30V DC, Resistive.

All specifications subject to change without notice.

GAUGE OPERATING TEMPERATURE

45°C

100°C

160°C

Range

0.1, 1, 2, 10, 100, 1000 Torr

1, 2, 10, 100, 1000

Accuracy

1

1000 Torr

100 Torr
20 Torr
10 Torr
2 Torr
1 Torr

<1 Torr

± 0.25% of reading
± 0.15% of reading
± 0.15% of reading
± 0.15% of reading
± 0.15% of reading
± 0.15% of reading
± 0.25% of reading

± 0.25% of reading
± 0.25% of reading
± 0.25% of reading
± 0.25% of reading
± 0.35% of reading
± 0.35% of reading

N/A

Measurement Range

4 Decades

Temperature Effects on Zero
(TCZO)

± 0.002% of FS/°C

Temperature Effects on Span
(TCFSO)

± 0.02% of FS/°C

Ambient Operating
Temperature

2

15°C to 35°C

15°C to 45°C

15°C to 35°C

Exposed Materials

Inconel

®

and/or AISI 316L Stainless Steel

Volume

17cc

Approximate Shipping Weight

1.40 lbs (726 grams)

Overpressure Limit

17 PSIA or 125% of Full Scale, whichever is greater

Power Input

Initial Warm-up -
Analog (Typical)

3,4

270mA @ 15 VDC (±5%)

620mA @±15 VDC (±5%) 620mA @±15 VDC (±5%)

Steady State -
Analog (Typical)

3

175mA @ 15 VDC (±5%)

350mA @±15 VDC (±5%) 550mA @±15 VDC (±5%)

Initial Warm-up -
DeviceNet

®

(Typical)

400mA @ 24 VDC

900mA @ 24 VDC

950mA @ 24 VDC

Steady State -
DeviceNet
(Typical)

300mA @ 24 VDC

600mA @ 24 VDC

600mA @ 24 VDC

Output

0 - 10 VDC (5k load) minimum

Relay Contact Rating

4

(available on 15 Pin Male D-
Sub Interface)

1A @ 30 VDC / 0.3A @ 125 VDC

Electrical Connector Type

Analog
RS-485
DeviceNet

9-pin D-sub male or 15-pin D-sub male
2.5 mm mini jack for RS-485 Communications Port
5-pin eurofast

®

(DeviceNet) w/9-pin D-sub Female (Analog)

Certification

Fully CE Compliant to EMC Directive 89/336/EEC

2.0

_DESCRIPTION

This manual is related to the following products: