P.17, 0 _description – Brooks Instrument XacTorr CMX160 User Manual
Page 23
p.17
Figure 17: DeviceNet Top Dimension
Figure 18: Zero Bias (Indy) 15-Pin
Table 4: CMX Top LED Indicators, Switches, and Connectors
ITEM
DESCRIPTION
DIAG
2.5mm Diagnostic Port - supports diagnostic functions via an RS-485 based communications
protocol.
ZERO
When pressed, the gauge will automatically adjust the pressure signal to zero volts.
MSD ADDRESS Most Significant address Digit (MAC ID).
ANALOG MODELS: For RS-485 Diagnostics Port.Valid addresses are 0 - 31.
DEVICENET MODELS: Valid addresses are 0 - 63 and P (Programmable).
LSD ADDRESS
Least Significant address Digit (MAC ID).
ANALOG MODELS: For RS-485 Diagnostics Port.Valid addresses are 0 - 31.
DEVICENET MODELS: Valid addresses are 0 - 63.
9-PIN D SUB
(MALE)
Power, analog pressure signal.
15-PIN D SUB
(MALE)
Power, analog pressure signal, and relay contacts.
READY/FAULT/
WARM UP
Temperature of heater indicator LED. Warm up = Amber, At Temperature = Green, Fault = Red
RELAY 1
Relay 1 status. When relay is energized, LED is lit Green.(Only on analog relay models)
RELAY 2
Relay 2 status. When relay is energized, LED is lit Green. (Only on analog relay models)
MOD STATUS
Module Status LED, per DeviceNet definition. (Only on DeviceNet models)
NET STATUS
Network Status LED, per DeviceNet definition. (Only on DeviceNet models)
RATE
Sets to DeviceNet data rate: 1 = 125kbps, 2 = 250 kbps, 5 = 500kbps, P = Program mode. (Only
on DeviceNet models)
5-PIN CONNEC-
TOR
“Micro-style” connector for DeviceNet power and communications.
77.4 mm
(3.05 inch)
O 81.0 mm
(3.19 inch)
2.0
_DESCRIPTION
77.4 mm
(3.05 inch)
O 81.0 mm
(3.19 inch)