p.28

5.0 MAINTENANCE AND TROUBLESHOOTING

5.1 CALIBRATION

The Brooks CMX has been manufactured, tested and is compliant to all
requirements of the applicable published specifications. All devices used for the
manufacture and test of these instruments have traceability to National Institute
of Standards (NIST) according to ISO 10012-1.

Digital CMs should be calibrated every 6-12 months depending on the
application and actual usage. The following factors may affect calibration and
may require calibration at more frequent intervals:

• Sensor diaphragm has become contaminated with process material

• Improper handling (dropping, misuse, etc.)

If the CMX is suspected to have calibration errors, return the unit to Brooks for
repair/calibration.

5.2 TROUBLESHOOTING

Table 10: Troubleshooting Guide

5.3 CLEANING

Cleaning is NOT recommended by anyone other than authorized Brooks service
personnel as this may cause damage to the device and may void the warranty.

Symptom

Possible Cause

Action

No output signal

No power

Check electrical connection.

Check cable continuity.

Verify power supply outputs.

Check for dents to the enclosure.

Fluctuating Zero

Temperature fluctuations

Check for varying temperature sources.

Power supply fluctuations

Check power supply regulation.

Fluctuating output signal

Electrical noise

Check power supply regulation.

Verify cable grounding and shields.

Pressure control system

Check pressure control system.

Excessive vibration

Reduce vibration.

Pressure reading too low
or too high

Zero offset

Verify electrical zero at zero pressure.

Power input wrong

Verify power voltage levels.

Grounds

Check signal and power grounds.

Positive output drift

Chamber outgassing

Re-pump system to base pressure.

Leak in chamber

Eliminate chamber leak.

Negative drift

Chamber pumping down

Re-zero sensor after pump-down.

Leaky sensor

Return device for repair.

Cannot zero CMX

Not at base pressure

Verify chamber is at base pressure.

Temperature not in range

Wait for Green READY LED to illuminate.

Improper ground

Check signal and power grounds.

Device malfunction

Return device for repair.

Orientation

Check or replace calibration position (see
note below).

NOTE: For 1 Torr and sub 1 Torr full scale ranges CMX, the specific mounting orientation for the device is specified for

optimal performance. Model number containing “V” in it stands for vertical mounting. Without letter “V” in model number
implies a horizontal mounting.