background image

p.26

4.4 DEVICE WARM-UP

When an CMX is installed and powered up on a system or tool, the signal output
will be variable for a period of time. Once the temperature profile of the device
reaches a steady-state condition and thermal saturation has occurred, the
variance in the signal output will be negligible.

Table 7: Recommended Ready To Use Times

Ready to Use: - Warm-up completed, flashing Amber LED now Green, signal
output change negligible and repeatable temperatures status signal level at 3.3
VDC.

4.5 HEATER/TEMPERATURE (WARM-UP/READY/FAULT) LED
FUNCTIONALITY

The sensor temperature is precisely controlled via the digital signal processor.
An intuitive tri-color LED is provided to indicate the three temperature states:
Warm-up, Ready to Use, and Fault. The following graph and table relate the
LED status to the actual sensor temperature.

Figure 25: Temperature vs. LED State

Table 8: LED Function Table

Device Temp

Range

Device Pressure

Range

From Cold Condition

(in hours)

45

o

C

100m Torr

5

1 Torr - 1000 Torr

3

100

o

C

100m Torr

5

1 Torr - 1000 Torr

3

160

o

C

100m Torr

5

1 Torr - 1000 Torr

3

LED Indicator

AMBER (Warming Up)

FLASHING AMBER (At temperature but not ready for use)

GREEN (Sensor at temperature and ready to use)

RED (Over Temperature) (Fault)

4.0

_OPERATION