Figure 22: 15-Pin “D” Connector Pinouts

3.7 DEVICENET CONNECTION

Set the DeviceNet baud rate selection switch (Located on top of the device) to
the desired baud rate. Set the address selection switches (Node Address) to the
desired address, noting the most significant digit and least significant digit
position. After these switches have been set, connect the DeviceNet power/
communication cable to the device. When power is applied, the unit will go
through an initialization procedure. The LED indicators will flash in sequence.
After this initial sequence, the Module Status LED should be green, the Network
Status LED should be flashing green (if network on-line, with no connection
established), and the Ready/Temperature status indicator should be:

• Amber – unit is warming up to the proper operating temperature.

PIN #

PIN ASSIGNMENT

1

Temperature in control relay NO

2

Pressure signal

3

Temperature in control relay common

4

Temperature in control relay NC

5

Power supply common

6

-15 VDC

7

+15 VDC

8

Relay 1 - Normally OPEN

9

Relay 1 - Contact common

10

Relay 1 - Normally CLOSED

11

Relay 2 - Normally OPEN

12

Signal common

13

Relay 2 - common

14

Relay 2 - Normally CLOSED

15

Chassis ground

PIN #

PIN ASSIGNMENT

1

Temperature in control relay NO

2

Pressure signal

3

Temperature in control relay common

4

Temperature in control relay NC

5

Power supply common

6

-15 VDC

7

+15 VDC

8

Relay 1 - Normally OPEN

9

Relay 1 - Contact common

10

Relay 1 - Normally CLOSED

11

Overheat Relay - Normally OPEN

12

Signal common

13

Overheat Relay - common

14

Overheat Relay - Normally CLOSED

15

Chassis ground

PIN #

PIN ASSIGNMENT

1

Temperature status

2

Output signal

3

Remote ZERO trigger

4

No connection

5

External power supply common

6

-15 VDC

7

+15 VDC

8

Relay 1 - Normally OPEN

9

Relay 1 - Contact common

10

Relay 1 - Normally CLOSED

11

Relay 2 - Normally OPEN

12

Signal common

13

Relay 2 - Contact common

14

Relay 2 - Normally CLOSED

15

Chassis ground

Connector Option 5 (CMX160 only)

Connector Option 2

Connector Option 4 (CMX160 only)

CMX

Cable

15-Pin D Sub Male

15-Pin D Sub Female

3.0

_INSTALLATION

This manual is related to the following products: