Brooks – Brooks Instrument GF81 User Manual
Page 43
4-5
Brooks
®
GF40/GF80/GF81 Devices
Installation and Operation Manual
X-TMF-GF40-GF80-Series-MFC-eng
Part Number: 541B161AAG
June, 2013
Section 4 Maintenance & Troubleshooting
Symptoms & Possible Causes
Corrective Action
2. Flow out of range (Continuted).
Is the device zero correct?
Zero the device according to zeroing
procedure in Section 2-16. Verify leak check rates
are OK.
3. No gas control; flow is at or
above maximum.
Is the gas pressure across the device
Verify that the pressure is correct for the gas and
too high?
range. If required, adjust inlet/outlet pressure to
achieve proper pressure reading.
Are system valves open, or is the
Use tool software to verify.
purge input activated?
Confirm that pin 12 is not connected to an active voltage
source.
Is the setpoint correct for the
Use tool software to verify.
required flow?
4. No gas flow above some setpoint.
NOTE: When the setpoint is
increased beyond this point,
the
GF40/GF80/GF81
Series signal remains at
some value lower than the set-point.
Is the gas inlet/outlet differential
Verify that the pressure is correct for
pressure sufficient?
the gas and range. If required, adjust
regulator to achieve proper pressure.
NOTE: If the pressure is too low,
the valve voltage to the
GF40/GF80/GF81
Series
will be at its maximum output. This condition
will cause internal GF valve heating and
inability to properly reach desired flow setpoints.
Is the
GF40/GF80/GF81
Series calibrated for the
Check
GF40/GF80/GF81
Series side label. Run a flow check
gas flow?
to verify. If flow is incorrect, replace the
GF40/GF80/GF81
Series with a unit that is calibrated properly.
Table 4-2 GF40/GF80/GF81 Series Troubleshooting Guide (Continued)