Save parameters into flash (x), Set parameters to factory default (z), Chapter – ALLIED Vision Technologies Goldeye P-032 SWIR User Manual
Page 80: Save, Parameters into flash (x)
Goldeye Technical Manual V2.5.0
80
Save parameters into flash (X)
Command
X=1 stores all parameter values into a system file within the flash
memory. After a restart of the camera they are automatically restored. However,
neither newly recorded correction images, nor modified set values (
J and K
for all valid
S) are stored and will therefore be lost after the camera is switched
off.
Set parameters to factory default (Z)
GigE Feature Name
Feature Visibility
Description
SaveParameterToCameraFlash
Expert
Stores all parameter values into a
system file within the camera‘s
flash memory.
(X=
Table 45: GigE feature: Save parameters into flash
Caution
The parameter Z is reserved for AVT internal use.
Unauthorized use of this parameter can render the camera
unusable.