B&K Precision 879B - Manual User Manual
Page 94

92
Frequency = 1 kHz
Range
Max
Display
Cx
Accuracy
DF(Dx
<0.5)
Measurement
Mode
1000 μF
1000.0 μF
3.7% + 3
digits
3.7% +
50 digits
Series
400 μF
399.99 μF
2% + 2
digits
2% + 50
digits
Series
40 μF
39.999 μF
0.7% + 2
digits
0.7% +
50 digits
Series
4000 nF
3999.9 nF
0.5% + 1
digit
0.5% +
50 digit
Series
400 nF
399.99 nF
0.5% + 2
digits
0.5% +
50 digits
---
40 nF
39.999 nF
0.5% + 2
digits
0.5% +
50 digits
---
4000 pF
3999.9 pF
0.7% + 2
digits
0.7% +
50 digits
Parallel
400 pF
399.9 pF
2.5% + 2
digits
2.5% +
50 digits
Parallel
Frequency = 10 kHz*
Range
Max
Display
Cx
Accuracy
DF(Dx
<0.5)
Measurement
Mode
100 μF
100.00 μF
3.9% + 5
digits
3.9% +
50 digits
Series
40 μF
39.999 μF
3.7% + 3
digits
3.7% +
50 digits
Series
4000 nF
3999.9 nF
0.7% + 2
digits
0.7% +
50 digits
Series