Recommendations – B&K Precision 879B - Manual User Manual

Page 54

background image

52

3. Select the measurement mode (i.e. Series or

Parallel)

4. Once all settings are configured, hold down

button for 2 seconds to enter into

calibration mode.

5. First, do open calibration. Follow instructions

under “Open Calibration” section.

6. Next, do short calibration. Follow instructions

under “Short Calibration” section.

7. Now, the meter should go back to normal

display and user may proceed to take

measurements of components.

Recommendations

To achieve optimal measurement results, both open

and short calibration should be done. It is highly

recommended to calibrate extremely high or low

values for L, C, R and Z before making precision

measurements.

Notes:

1. If test frequency is changed, calibration

should be done again before making precision

measurements. Once calibration is done on a

selected test frequency, calibration data will

remain until power off.