Supplemental information, Selecting test frequency, Capacitance – B&K Precision 879B - Manual User Manual

Page 82

background image

80

SUPPLEMENTAL

INFORMATION

This section provides supplemental information for

user consideration when operating the LCR meters.

Some recommendations and explanations are

provided to help aid in the use of some functions and

features, in which can help the user gain optimal and

accurate measurement results.

Selecting Test Frequency

Test frequency can greatly affect the results of

measurement reading, especially when measuring

inductors and capacitors. This section provides some

recommendations and suggestions to consider.

Capacitance

When measuring capacitance selecting, the right

frequency is important in obtaining the most accurate

measurement results. Generally, a 1 kHz test

frequency is used to measure capacitors that are 0.01

µF or smaller. For capacitors that are 10 µF or larger,

a lower frequency of 120 Hz is used. Following this

trend, high test frequencies are best for testing very

low capacitance components. For large capacitance

components, low frequency would be optimal. For