Storable settings, Configure and save power-up state – B&K Precision 879B - Manual User Manual
Page 45
43
In the utility menu, when the primary display shows
“
PuP”, there are two settings selectable and shown on
the secondary display. “
PrE” and “SEt”.
Default Setting: PrE
Storable Settings
•
Primary function mode (i.e. L/C/R)
•
Test frequency
•
Secondary function mode (i.e. D/Q)
•
Tolerance mode state
•
Reference value for Tolerance mode
•
Relative mode state
•
Reference value for Relative mode
Configure and Save Power-up State
Follow the below procedure to setup and store a
power-up state into internal memory.
1. Before entering into the utility menu, configure
all the settings and parameters desired for
power up state. Do this by turning on any
modes and setting values as desired. (Only
the settings listed above in “Storage Settings”
are stored). If the meter is currently in the
utility menu, exit first and setup the desired