out of accuracy or erroneous. Series mode is use

because large capacitors often have higher

dissipation factor and lower internal resistance.

Inductance

For most inductance measurement, selecting

series

mode is the best. Hence, the meter defaults to this

mode when selecting inductance mode. This is

because in this mode, accurate Q (quality factor)

reading can be obtained from reading low Q inductors

and ohmic losses are significant.

Though in some cases,

parallel mode would be

preferred. For example, iron core inductors operating

at higher frequencies where hysteresis and eddy

currents become significant would require

measurement in parallel mode for optimal results.

Accuracy Discrepancies

In some special cases, inaccuracies may occur in the

measurement of capacitive, inductive, and resistive

components.