Frequency – Atec Agilent-N5182A User Manual
Page 4
4
Range
Option 503
100 kHz to 3 GHz
Option 506
100 kHz to 6 GHz
Minimum frequency
100 kHz
1
Resolution
0.01 Hz
Phase offset
Adjustable in nominal 0.1 ° increments
Frequency bands
2
Band
Frequency range
N
1
100 kHz to < 250 MHz
1
2
250 to < 375 MHz
0.25
3
375 to < 750 MHz
0.5
4
750 to < 1500 MHz
1
5
1500 to < 3000.001 MHz
2
6
3000.001 to 6000 MHz
4
Switching speed
3, 4, 6
Type
Standard
Option UNZ
5
Option UNZ
5
(typical)
Digital modulation off
SCPI mode
≤ 5 ms (typ)
≤ 1.15 ms
≤ 950 µs
List/Step sweep mode ≤ 5 ms (typ)
≤ 900 µs
≤ 700 µs
Digital modulation on
SCPI mode
≤ 5 ms (typ)
≤ 1.15 ms
≤ 1.05 ms
List/Step sweep mode ≤ 5 ms (typ)
≤ 900 µs
≤ 800 µs
1. Performance below 250 kHz is unspecified except as indicated, for units with serial numbers
ending with 4742xxxx or greater. For units with lower serial numbers refer to the Archive
Section at end of this document.
2. N is a factor used to help define certain specifications within the document.
3. Time from receipt of SCPI command or trigger signal to within 0.1 ppm of final frequency
or within 100 Hz, whichever is greater, and amplitude settled to within 0.2 dB.
4. Additional time may be required for the amplitude to settle within 0.2 dB when switching
to or from frequencies < 500 kHz.
5. Specifications apply when status register updates are off.
6. With Internal Channel Corrections on, the frequency switching speed is < 1 ms (measured) for
list mode and SCPI mode cached frequency points. For the initial frequency point in SCPI mode
the time is < 75 ms (measured). The instrument will automatically cache the most recently
used 256 frequencies. There is no speed degradation for amplitude-only changes. Internal
Channel Correction applies to FW A.01.60 or greater with Option N5162/82AK-R2C.