Atec Agilent-N5182A User Manual
Page 29
29
Weight
≤ 12.5 kg (27.5 lb.) net, ≤ 27.2 kg (60 lb.) shipping
dimensions
88 mm H x 426 mm W x 432 mm L
[3.5 in H x 16.8 in W x 17 in L]
Recommended
calibration cycle
36 months. Agilent is committed to providing you with the
lowest total cost to own and operate equipment. In support
of this commitment, Agilent has verified that the stability of
this product's architecture justifies a longer calibration inte
val of 3 years.
ISO compliant
The Agilent N5182A MXG is manufactured in an ISO-9001
registered facility in concurrence with Agilent Technologies’
commitment to quality.
Front panel connectors
1
RF output
2
Outputs the RF signal via a precision N type female connector.
Maximum reverse power is 2 W, 50 VDC.
I and Q inputs
2
Accepts “in-phase” and “quadrature” input signals for I/Q
modulation. Nominal input impedance is 50 Ω. Damage
levels are 1 Vrms and 5 Vpeak.
USB 2.0
Used with a memory stick for transferring instrument states,
licenses and other files into or out of the instrument. Also
used with U2000 Series USB average power sensors.
For a current list of supported memory sticks, visit
www.agilent.com/find/MXG, click on Technical Support,
and refer to FAQs: Waveform Downloads and Storage.
Rear panel connectors
1
RF output
Outputs the RF signal via a precision N type female connector.
(Option 1EM or N5162A)
I and Q inputs
Accepts “in-phase” and “quadrature” input signals for I/Q
(Option 1EM or N5162A) modulation. SMB connector, nominal input impedance is
50 Ω. Damage levels are 1 Vrms and 5 Vpeak. Option 1EM and
N5162A units will come with 2 SMB to BNC adapters.
I and Q outputs
Outputs the analog I/Q modulation signals from the internal
baseband generator. Nominal output impedance 50 Ω,
DC coupled. Damage levels ±2 V.
I– and Q–
outputs
Outputs the complement of the I and Q signals for differential
(Option 1EL)
applications. Nominal output impedance is 50 Ω, DC-coupled.
Damage levels are ±2 V.
EXT Clk
Reserved for future use.
Event 1
This connector outputs the programmable timing signal
generated by marker 1. The marker signal can also be routed
internally to control the RF blanking and ALC hold functions.
This signal is also available on the AUX I/O connector.
This output is TTL and 3.3 V CMOS compatible. Damage
levels are > +8 V and < –4 V.
Pattern trigger
Accepts signal to trigger internal pattern generator to start
single pattern output, for use with the internal baseband
generator (Option 651, 652, 654). Accepts CMOS
3
signal with
minimum pulse width of 100ns. Female BNC; Damage levels
are > +8 V and < –4 V.
Sweep out
Generates output voltage, 0 to +10 V when the signal
generator is sweeping. This output can also be programmed to
indicate when the source is settled or output pulse video and
is TTL and CMOS compatible in this mode. Output
impedance < 1 Ω, can drive 2 kΩ. Damage levels are ±15 V.
1. All connectors are BNC unless otherwise
noted.
2. All N5162A MXG ATE connectors located
on rear panel.
3. Rear panel inputs and outputs are 3.3 V
CMOS, unless indicated otherwise. CMOS
inputs will accept 5 V CMOS, 3 V CMOS,
or TTL voltage levels.