General characteristics – Atec Agilent-N5182A User Manual
Page 28
28
Remote programming
Interfaces
GPIB
IEEE-488.2, 1987 with listen and talk
LAN
100BaseT LAN interface,
LXI class C compliant
2
USB
Version 2.0
Control languages
SCPI
Version 1997.0
Compatibility languages supporting 100% of commonly used commands
1
Agilent Technologies
E4438C, E4428C, E442xB, E443xB, E8241A,
E8244A, E8251A, E8254A, E8247C, E8257C/D,
E8267C/D, 8648 series, 8656B, E8663B, 8657A/B
Aeroflex Incorporated
3410 series
Rohde & Schwarz
SMU200A, SMJ100A, SMATE200A, SMIQ,
SML, SMV
Power requirements
100 or 120 VAC, 50 or 60 Hz, 400 Hz
220 or 240 VAC, 50 or 60 Hz
250 W maximum
Operating temperature range
0 to 55 °C
Storage temperature range
–40 to 70 °C
Operating and storage altitude
Up to 15,000 feet
Environmental stress
Samples of this product have been type tested in
accordance with the Agilent Environmental Test
Manual and verified to be robust against the
environmental stresses of Storage, Transportation
and End-use; those stresses include but are not
limited to temperature, humidity, shock, vibration,
altitude and power line conditions. Test Methods
are aligned with IEC 60068-2 and levels are similar
to MIL-PRF-28800F Class 3.
Safety
Complies with European Low Voltage Directive
73/23/EEC, amended by 93/68/EEC
· IEC/EN 61010-1
· Canada: CSA C22.2 No. 61010-1
· USA: UL 61010-1
EMC
Complies with European EMC Directive
89/336/EEC, amended by 93/68/EEC
· IEC/EN 61326
· CISPR Pub 11 Group 1, class A
· AS/NZS CISPR 11:2002
· ICES/NMB-001
Memory
Memory is shared by instrument states, user
data files, sweep list files, waveform sequences,
and other files. There are 4 GB of flash memory
available in the N5182A MXG. Depending on
how the memory is utilized, a maximum of 1000
instrument states can be saved.
Security (Option 006)
Memory sanitizing, memory sanitizing on power
on, and display blanking
Self test
Internal diagnostic routines test most modules in
a preset condition. For each module, if its node
voltages are within acceptable limits, the
module “passes” the test.
1. Firmware version A.01.10 and later.
2. LXI class B compliant with Option ALB.
General Characteristics