Atec Agilent-N5182A User Manual
Page 33
33
SWR
≤ 2.1 GHz
1.4:1 (typ)
(for serial number
> 2.1 GHz to 4 GHz
1.5:1 (typ)
prefix 4742xxxx)
> 4.0 GHz 5.6 GHz
1.7:1 (typ)
> 5.6 GHz to 6 GHz
2.0:1 (typ)
Maximum
Max DC voltage
50 VDC (nom)
reverse power
250 kHz to 6 GHz
2 W (nom)
SWR
≤ 1.4 GHz
1.7:1 (typ)
(for serial number
> 1.4 GHz to 4 GHz
2.3:1 (typ)
prefixes lower than
> 4.0 GHz 5.0 GHz
2.4:1 (typ)
4742xxxx)
> 5.0 GHz to 6 GHz
2.2:1 (typ)
Maximum
Max DC voltage
50 VDC (nom)
reverse power
50 kHz to 6 GHz
2 W (nom)
Absolute level accuracy in CW mode
1
[ALC on]
(for serial number prefix 4742xxxx)
Standard
Option 1EQ
+7
2
to -60 dBm < -60 to -110 dBm
< -110 to -127 dBm
100 kHz to 250 kHz
±0.6 dB
±1.0 dB
---
> 250 kHz to 1 MHz
±0.6 dB
±0.7 dB
±1.7 dB
> 1 MHz to 1 GHz
±0.6 dB
±0.7 dB
±1.0 dB
> 1 GHz to 3 GHz
±0.7 dB
±0.9 dB
±1.4 dB
> 3 GHz to 4 GHz
±0.8 dB
±0.9 dB
±1.0 dB
> 4 GHz to 6 GHz
±0.8 dB
±1.1 dB
±1.3 dB
Absolute level accuracy in CW mode
1
[ALC on]
(for serial number prefixes lower than 4742xxxx)
Standard
Option 1EQ
+7 to -60 dBm
< -60 to -110 dBm
< -110 to -127 dBm
250 kHz to 1 MHz
±0.6 dB
±0.7 dB
±1.7 dB
> 1 MHz to 1 GHz
±0.6 dB
±0.7 dB
±1.0 dB
> 1 GHz to 3 GHz
±0.7 dB
±0.9 dB
±1.4 dB
> 3 GHz to 4 GHz
±0.8 dB
±0.9 dB
±1.0 dB
> 4 GHz to 6 GHz
±0.8 dB
±1.1 dB
±1.3 dB
1. Quoted specifications between 20 and
30 °C. For temperatures outside of this
range, absolute level accuracy degrades
by 0.01 dB/ °C for frequencies ≤ 4.5 GHz
and 0.02 dB/ °C for frequencies
> 4.5 GHz.
2. Level accuracy specified to +7 dBm or
maximum specified output power, which-
ever is lower.
Maximum available output power
Frequency (MHz)
Max output power (dBm)
0
1000
2000
3000
4000
5000
6000
18
17
16
15
14
13
12
11
10