Definitions – Atec Agilent-N5182A User Manual

Page 3

background image

3

Specification (spec):

Represents warranted performance of a calibrated instrument

that has been stored for a minimum of 2 hours within the operating temperature range of

0 to 55 °C, unless otherwise stated, and after a 45 minute warm-up period. The specifi-

cations include measurement uncertainty. Data represented in this document are

specifications unless otherwise noted.

Typical (typ):

Represents characteristic performance, which 80% of the instruments

manufactured will meet. This data is not warranted, does not include measurement

uncertainty, and is valid only at room temperature (approximately 25 °C).

Nominal (nom):

The expected mean or average performance, or an attribute whose

performance is by design, such as the 50 Ω connector. This data is not warranted and is

measured at room temperature (approximately 25 °C).

Measured (meas):

An attribute measured during the design phase for purposes of

communicating expected performance, such as amplitude drift vs. time. This data is not

warranted and is measured at room temperature (approximately 25 °C).

Note: All graphs contain measured data from several units at room temperature unless

otherwise noted.

Definitions