Pulse train (option 320), External modulation inputs, Simultaneous modulation – Atec Agilent-N5182A User Manual
Page 16

16
Internal pulse generator (included with Option UNU or Option UNW)
Modes
Free-run, square, triggered, adjustable doublet,
trigger doublet, gated, and external pulse
Square wave rate
0.1 Hz to 10 MHz, 0.1 Hz resolution (nom)
Pulse period (UNU)
500 ns to 42 seconds (nom)
Pulse width (UNU)
500 ns to pulse period – 10 ns (nom)
Pulse period (UNW)
30 ns to 42 seconds (nom)
Pulse width (UNW)
20 ns to pulse period – 10 ns (nom)
Resolution
10 ns
Adjustable trigger delay:
–pulse period + 10 ns to pulse period
to pulse width –10 ns
Settable delay
Free run
–3.99 to 3.97 µs
Triggered
0 to 40 s
Resolution
[delay, width, period]
10 ns (nom)
Pulse doublets
1st pulse delay
(relative to sync out) 0 to 42 s – pulse width – 10 ns
1st pulse width
500 ns to 42 s – delay – 10 ns
2nd pulse delay
(relative to pulse 1) 0 to 42 s – (delay1 + width2) – 10 ns
2nd pulse width
20 ns to 42 s – (delay1 + delay2) – 10 ns
Pulse train (Option 320)
Number of pulse patterns: 2047
On/off time range (UNU): 500 ns to 42 sec
On/off time range (UNW): 20 ns to 42 sec
External modulation inputs
1
Modulation types
FM, AM, phase mod, pulse mod
Input impedance
50 Ω (nom)
Simultaneous modulation
2
All modulation types (FM, AM,
φ
M and pulse modulation) may be simultaneously
enabled except: FM and phase modulation can not be combined; two modulation
types can not be simultaneously generated using the same modulation source.
For example the baseband generator, AM, and FM can run concurrently and all will
modulate the output RF. This is useful for simulating signal impairments.
1. Option UNT required for FM, AM, and phase mod inputs. Option UNU or UNW required for
pulse modulation inputs.
2. If AM or pulse modulation are on then phase and FM specifications do not apply.