33 

Chapter 6: Specifications

Model Number:
CTE-1200 AT; CTE-2200 AT; CTE-3200 AT

Power Requirements:
One 9-volt high capacity/heavy duty lithium

battery or rechargeable battery pack accessory

Applications:
Tests individual lead acid cells or monoblocs

(6, 12 volts) in any common configuration

Environmental Operating Range:
0 to +40°C, 95% relative humidity,

non-condensing

Voltage:
1.0–15.0 volts DC

Storage Temperature:
–20 to 82°C

Conductance:
100–9,999 siemens

Over Voltage Protection:

Fused protection to 60 volts DC

Reverse polarity protected

Test Data Storage:
Up to 4 strings per site

Housing Material:
Acid-resistant ABS plastic

Accuracy:
+ 2% across test range

Analyzer Dimensions:
9 in x 4 in x 2.5 in
230 mm x 102 mm x 65 mm

Voltmeter Resolution:
10 mV DC

Case Dimensions:
19 in x 15.5 in x 5 in
750 mm x 610 mm x 200 mm

User Programmable Functions:

Low voltage alarm setting

Low conductance warning

Low conductance failure

Day/date/time formats (USA/international)

Test mode (auto start)

Analyzer Weight:
1 lb / 500 gm

Shipping Weight: CTE-3200 AT Kit
9.5 lb / 4 kg

Calibration:
Auto-calibration prior to every test; no future

calibration required

Special Features:

Impact-resistance tested

Connection interfaces tested for durability

and endurance

No-Ox grease petroleum product resistance

Test Cable Options:

Dual contact clamps

Dual contact probes

Custom cables by quotation