beautypg.com

N-x144_iom_18, 1 – starting operation – Cla-Val X144 Technical Manual User Manual

Page 18

background image

SECTION 3: X144 e-FlowMeter Operation

3.1 – Starting Operation

Follow this procedure when starting the installation for the first time or after shutting the system down for

maintenance of the Control Valve.

1) Check all connections to power supply and other User Devices (data loggers, telemetry, etc.)

2) Ensure all wiring is properly connected.

3) Open shut-off valves and allow water into the system.

4) Check the system for water leaks, at both the tapered threads, as well as the straight threads

with knurled lock.

5) Check surrounding piping and auxiliary equipment for leaks.

6) Check X144 e-FlowMeter output and verify proper operation at current flow condition

(per table 1):

Table 1: Minimum and Maximum Flow values for X144 e-FlowMeter

15

Log onto www.cla-val.com

and click on the X144 e-FlowMeter

icon to learn more

Line Size

inches (mm)

*2"

(50mm)

(100-49 Body)

2-1/2"

(65mm)

3"

(80mm)

4"

(100mm)

6"

(150mm)

8"

(200mm)

10"

(250mm)

12"

(300mm)

16"

(400mm)

Minimum Flow

(GPM)

10

10

15

25

60

100

160

230

360

Maximum Flow

(GPM)

210

300

460

800

1800

3150

4950

7000

11000

Minimum Flow

(l/S)

0.65

0.65

1.0

1.5

3.8

6.3

10.0

15

23

Maximum Flow

(l/S)

13.3

18.9

29

50

110

200

310

440

700

*2" X144 e-FlowMeter may be installed on new valves only
Note:

Consult Factory for Angle Pattern Applications