Power Supply
Power Supply
82
BA31114e5 07/2006
BA31114e5 07/2006
83
OxiTop
®
-C measuring head
Supply batteries
Batteries: 2 x lithium batteries, CR2430, e.g. WTW type Batt/OxiTop
,
order no. 209 012
These batteries ensure the energy supply of the OxiTop
®
-C measuring head.
Run time: typically 2 years
The OxiTop
®
-C measuring head does not require any batteries for data backup.
Supply batteries status messages
- are given on starting the measuring head via the OxiTop
®
OC100 controller by
display messages and a signal tone
- can be called up in the menu “GLP/Tools - Check -
info”.
Display message
Note / Meaning
Batteries LoBat !
start
The instrument can still be safely operated within the
specifications.
The running reserve of the instrument is still sufficient for
the measurement to be started.
Please obtain new batteries and replace the old ones!
Battery/ies empty!
- New
- Stop
start
The instrument can no longer be used.
The supply batteries must be replaced by new ones.
It is possible to start a new measuring head or to
complete a sample start.
Changing the supply battery
• Change the battery after the measurement has been terminated.
- Press in the snap-action holder.
- Open the upper case.
- Remove the batteries.
- Insert new batteries (2 x lithium batteries, CR2430, e.g. WTW type
Batt/OxiTop
, order no. 209 012). Ensure the batteries are inserted the right
way round!
- Insert the upper case with the lug in the locking pin (lower case).
Caution! Do not crush the cable connection!
- Close the upper case (let the hook snap into place).
Check if the battery change was successful:
“Batteries: OK“ must appear in “GLP/Tools - Check -
info”.
The batteries have no effect on the data integrity of the OxiTop
®
-C
Disposing of the batteries
Properly dispose of used batteries at a battery collection point.