Xylem OxiTopIS OC100 User Manual
Page 37
Operating Mode: Standard BOD
Operating Mode: Standard BOD
36
BA31114e5 07/2006
BA31114e5 07/2006
37
According to the sample selected, one of the two following menus
appears. The header line contains the:
- sample number
- Id number (e.g. I001)
- BOD type and run time
of the selected sample.
Finished sample
Current sample
Show sample
This function is used to find the measuring heads or samples. Sample labeling is not
necessary.
Starting point: sample
management .
Use
to select a sample.
Point the controller at the measuring
heads:
40 cm to 100 cm
The controller transmits the call
up of the selected sample. The
allocated measuring head flashes
for approx. 5 seconds.
After the message, the controller automatically returns to the
previous menu.
Erase data of finished samples
This function erases the data of finished samples from the sample management of
the controller. You can only erase a sample if it is finished; in non-finished samples,
the menu item “Erase sample“ does not appear.
Starting point: sample
management .
Use
to select a finished sample.
Use
to select the “Erase sample“
submenu.