Xylem OxiTopIS OC100 User Manual
Page 43

Operating Mode: Standard BOD
Operating Mode: Standard BOD
42
BA31114e5 07/2006
BA31114e5 07/2006
43
Call up data - Stop
If a measuring head is missing or
defective and the controller
cannot completely call up the
sample as a result, the menu
item “Stop“ is provided to stop
the data transfer.
After confirmation, the display
message shown here appears.
Three displays are then possible:
Case 1
The sample is still running and
individual measuring heads of the
sample are not attainable.
Case 2
The run time of the sample has
ended and the finished, complete
data of the attainable measuring
heads has been transferred.
then
Erase:
Removes missing measuring
heads from the data stock. The
completed sample is given the
“finished” status.
Back:
Jumps back to the previous
menu.
Abort using a function key: The
stopped sample is not declared
as “finished”.
Case 3:
No measuring head of a sample
is attainable.
then
Erase:
Removes a sample from the data
stock.
Back:
Jumps back to the previous
menu.
“Show
”:
As in “Show sample“ but selectively for missing measuring heads
(Identification option if only the sender of the measuring head is
defective. Otherwise, see the chapter “What to do if...?”)
“Missing
”:
Only appears when the run time of the sample has expired. It is used
to set the sample to the “finished” status (ticked) if all the other
measuring heads are already “finished”: Erases the missing
measuring heads from the sample management.