Print – Xylem OxiTopIS OC100 User Manual
Page 50

Operating Mode: Standard BOD
Operating Mode: Standard BOD
50
BA31114e5 07/2006
BA31114e5 07/2006
51
Measured values outside the measuring range
In the following cases, “undef.“ (undefined) is displayed instead of the measured
value or mean value during evaluation:
- A measured curve exceeding the measuring range at any point of its path.
- A measured curve undercutting the measuring range at its end point.
After the exclusion of the defective curve (see chapter “Exclude curve“), the controller
displays the mean value again.
Sample displays:
A measured value exceeding the
measuring range (Overflow).
A measured value undercuts the
measuring range (Underflow).
A measured value temporarily
leaves the measuring range.
Switch on the TD100 IR printer.
Point the controller at the printer:
10 cm to 100 cm
The controller outputs the report data of the selected sample to
the IR interface.
Printout of
- sample statistics (mean value, SD, n)
- sample information (type, running time, measuring range, start
date and start time, end date, identification no.)
- curve + BOD final value
- with GLP On: additional information on the system status
and list of the series nos. of the measuring heads allocated to
the sample
- If AutoTemp is switched on: note “AutoTemp on“
- Report printouts:
- current settings
- start information
-
info
- controller info
From the sample management:
From the evaluation (display of
all the curves):