Call up data
This function is used to call up the data of individual measuring heads. To call up the
data of all measuring heads, see the chapter “Call up all data“.
Use
to select a sample whose
- run time is not yet completed
(bar not yet full)
- data after complete
measurement that has not yet
been called up
(bar full, no tick)
Jumps to the submenu of the
selected sample.
Preselected: “Call up data“
Point the controller at the measuring
heads:
40 cm to 100 cm
Call up the measuring head
allocated to the sample in the
scanner mode:
The controller
- stores the data obtained
- updates the lists of the
sample management
The controller confirms the call
up performed.
The controller then returns
automatically to the submenu.
If the measuring head does not respond, the following display appears:
The controller continues to search for the measuring head in the scanner mode.
Without pressing any further keys, the data call up can be continued at other
locations (e.g. other stirring platforms or other thermal cabinets). After approx. 7
seconds, the scanner mode automatically breaks off and the following display
appears:
Use
to restart the
call up of the data (see above).
Return to the main menu,
“Sample management“.
Samples that were called up and
completed appear with a tick next
to them.