Xylem OxiTopIS OC100 User Manual
Page 33

Operating Mode: Standard BOD
Operating Mode: Standard BOD
32
BA31114e5 07/2006
BA31114e5 07/2006
33
The start was stopped.
Use
to confirm
“Continue
start” and hold the
controller to a measuring head
(see above).
Select “Stop
start” to return to
entry level menu.
If an attempt is made to start a measuring head that was already
started:
Displays “Measuring head
already used!”
Use
to confirm “New
” and hold the controller to a
free measuring head (see
above).
Select “Stop
start” to return to
entry level menu.
In the delivery state, the controller automatically makes space when
“Start
” is selected if the measured value memory is full. To do this,
it deletes the oldest finished sample (if a finished sample is
available). You can change this setting in “manual erase” (see the
chapter “GLP/Tools - Settings”).
Call up all data
This function is used to call up the data of all measuring heads.
To call up the data of individual measuring heads: see the chapter “ Sample
management”.
Entry menu “Communication with
the measuring heads“.
Use
to select “Call up all data”.
Point the controller at the measuring
heads:
40 cm to 100 cm
Queries the data of all active
measuring heads in the scanner
mode.
The controller stores the data
obtained and updates the list of
the sample management.
Duration of a pass: approx. 3
seconds for 12 measuring heads
in a stirring system.
If the measuring heads do not all respond in the 1st pass of the data query, the
controller searches for the missing measuring heads in the scanner mode for
approximately a further 7 seconds. Without having to press another key, you can
continue the data queries at other locations (e.g. other stirring platforms or other
thermal cabinets). Approximately 7 seconds after the last call up was performed, the
scanner mode breaks off automatically and the following display appears:
“Continue transfer“ (preselected):
A new pass is made only for
measuring heads that have not
yet responded.
Stop: The instrument returns to
the entry menu.
For information on searching for
missing measuring heads, see