Oxitop, Oc100 controller – Xylem OxiTopIS OC100 User Manual

Page 79

background image

Power Supply

Power Supply

78

BA31114e5 07/2006

BA31114e5 07/2006

79

The OxiTop

®

-OC100 and OxiTop

®

-C instruments are battery-powered.

To ensure reliable operation, both instruments have a 2-stage battery status monitor
1st level = warning level : Batteries LoBat !
2nd level = error level : Batteries empty !

OxiTop

®

-

OC100 controller


Economy circuit (automatic switch off)
The instrument switches off automatically following the last key actuation after expiry
of the set switch-off interval.


Switch-off interval:
• Routine BOD operating mode: preset to a fixed period of 5 min.

• Standard BOD operating mode: delivery state of 5 min.,

settings of 5...15 min. possible


Supply batteries


Batteries: 3 pcs, alkaline (alkaline manganese), size: Mignon, AA, AM3, LR6
These batteries ensure the energy supply of the OxiTop

®

OC100.

Run time: > 100 h (approx. 1000 start-ups in normal use)

Supply batteries status signals
are given on switching on the instrument by display messages and a signal tone
or can be called up under “GLP/Tools - Check - Controller info”.

Display message

Note / Meaning


Supply
Battery/ies LoBat !

Please change !

The warning appears for approx. 3 seconds.
The instrument then continues to run normally.
The instrument can still be safely operated within the
specifications.
When the message first appears, there is still a
running reserve available.
Please obtain new batteries and replace the old ones
!


Supply
Battery/ies empty!

Please change !

The message appears for approx. 3 seconds.
The instrument then switches itself off.
The instrument can no longer be used.
The supply batteries must be replaced by new ones.

Changing supply batteries:

• Switch off the OxiTop

®

-OC100 controller.

• Loosen the 4 screws underneath the housing using a Phillips screwdriver (see

figure

).

• Place the controller on the lower case.

• Remove the upper case and put it down to the right next to the lower case with

the display downwards (see figure

).

• Remove the supply battery holder from the fixing in the lower case and turn it

around.

• Remove the empty supply batteries.

• Insert new supply batteries - 3 pieces, alkaline (alkaline manganese),

size: Mignon, AA, AM3, LR6).
Ensure that they are the right way round!
(The poles are marked in the supply battery holder)
Always replace the complete set of batteries.

• Turn the supply battery holder around again and place it in the fixing in the lower

case.

• Set the upper case on top of the lower case, turn the controller and tighten the

housing screws using the screwdriver (see figure

).

• Switch on the controller. The battery change was successful if no error message

appears concerning the supply batteries.

• It is recommended to use only brand name batteries of the type specified as

“Alkaline“. Using other types of battery can affect reliable functioning.

• The supply batteries have no influence on data integrity.