Advanced instruments inc – Analytical Industries GPR-1600 Series Trace PPB Oxygen Analyzer User Manual
Page 43

Advanced Instruments Inc.
 
 
 
43
The following display appears: 
 
Sample
 080.00 PPM 
 
 
 Press UP or DOWN keys to change values 
 
 Select ENTER to save, ESC to return 
 
 
Auto Range 0 to 10 PPM 
85⁰F 100Kpa 12/31/07 12:00:00 
 
Press the ENTER key to move the cursor (underscore) to the right to the digit to be changed. 
Press the UP or DOWN key until the desired number appears in digit field. 
Repeat as necessary to enter the numerical span value. 
Press the ENTER key to accept and save the span gas value. 
 
The system then initiates the SPAN CALIBRATION function and the following menu appears: 
 
CALIBRATION
IN PROGRESS . . .
 
 
 SPAN VALUE ACTUAL O2 VALUE 
 
 
After the oxygen reading has stabilized, press ENTER to complete the Span Calibration. 
 
If the oxygen reading is outside +50% of the span gas value, the system displays the message “CALIBRATION FAILED” and 
returns to the “Sample” mode. 
 
 
CALIBRATION
FAILED
 
 
 
 
 
 
 
 