00 ppm, Advanced instruments inc, Alarms – Analytical Industries GPR-1600 Series Trace PPB Oxygen Analyzer User Manual
Page 28

Advanced Instruments Inc.
 
 
 
28
If the oxygen reading exceeds 99.9% of the full scale fixed range manually selected, the system displays the following: 
 
* MAIN MENU
 Sample 
 Span 
 Zero 
 Alarm 
 System 
 Standby 
 
Sample
5.00 PPM
OVER RANGE
Manual Range 
85⁰F 100Kpa 
0 to 10 PPM 
12/31/07 12:00:00 
 
Alarms 
The analyzer is configured with two user adjustable threshold type alarm relays that can be configured in the field from the 
ALARM option on the MAIN MENU as follows: 
 
¾ Establish independent set points 
¾ Either Hi or Lo 
¾ Either On or Off (enabled or disabled) 
¾ Both temporarily defeated using a user entered ‘timeout’ period (normally minutes) 
 
The alarm set point represents a value. When the oxygen reading exceeds (high alarm) or falls below (low alarm) the alarm set 
point, the relay is activated and the LCD displays the alarm condition. 
 
When activated the alarms trigger SPDT Form C non-latching relays @ 5A, 30VDC or 240VAC resistive. To prevent chattering of 
the relays, a 2% hysteresis is added to the alarm set point. This means that the alarm will remain active until the oxygen 
reading has fallen 2% below the alarm set point (high alarm) or risen 2% above the alarm set point (low alarm) after the alarm 
was activated. The timeout feature is useful while replacing the oxygen sensor or during calibration when the oxygen reading 
might well rise above the alarm set point and trigger a false alarm. 
 
Note: When making connections the user must decide whether to configure/connect Alarm 1 and Alarm 2 in failsafe mode 
(Normally Open – NO – where the alarm relay de-energizes and closes in an alarm condition) or non-failsafe mode (Normally 
Closed – NC – where alarm relay energizes and opens in an alarm condition). 
 
Procedure: Advance the cursor (*) to the “Alarm” option and press the green ENTER key to accept the selection. 
 
MAIN MENU
 Sample 
 Span 
 Zero 
* Alarm 
 System 
 Standby 
 
Sample
5.00 PPM
Auto Range 
85⁰F 100Kpa 
0 to 10 PPM 
12/31/07 12:00:00 
 
The following menu appears: 
 