00 ppm, Advanced instruments inc – Analytical Industries GPR-1600 Series Trace PPB Oxygen Analyzer User Manual

Page 38

background image

Advanced Instruments Inc.



38

ZERO Sample

* Factory Default
Calibrate





Auto Range 0 to 10 PPM
85⁰F 100Kpa 12/31/07 12:00:00


Advance the cursor (*) to the Factory Default Zero option and press ENTER.
Within seconds the system returns to the MAIN MENU in the “Sample” mode.

MAIN MENU

Sample
Span
* Zero
Alarm
System
Standby

Sample

5.00 PPM

Auto Range
85⁰F 100Kpa

0 to 10 PPM
12/31/07 12:00:00


Advance the cursor (*) to the “Zero” option and press the green ENTER key to accept the selection.
The following menu appears:

ZERO Sample

Factory Default
* Calibrate




Auto Range 0 to 10 PPM
85⁰F 100Kpa 12/31/07 12:00:00


Advance the cursor (*) to the “Manual Zero” option and press the green ENTER key to accept the selection.

The following menu appears: