42

MAIN MENU

Sample
* Span
Zero
Alarm
System
Standby

Sample

20.9 %

Auto Range
85⁰F 100Kpa

0 to 10 PPM
12/31/07 12:00:00


The following menu appears:

SPAN Sample

Factory Default
* Calibrate





Auto Range 0 to 10 PPM
85⁰F 100Kpa 12/31/07 12:00:00


Advance the cursor (*) to the “Manual Span” option and press the green ENTER key to accept the selection.
The following menu appears:

SPAN GAS Standby



* Enter as PPM
Enter as PPB
Enter as %


Auto Range 0 to 10 PPM
85⁰F 100Kpa 12/31/07 12:00:00


Enter span gas units appropriate to the numerical span gas value and press ENTER key.
Note: Select “Enter as %” for air calibration.