00 ppm, Advanced instruments inc – Analytical Industries GPR-1600 Series Trace PPB Oxygen Analyzer User Manual
Page 40

Advanced Instruments Inc.
 
 
 
40
Procedure – Span Calibration 
The analyzer is online in the Auto Range mode as described above. 
Review the Intervals, Online Recovery Time, Span Gas Selection, Type of Sensor, Span Calibration Adjustment and Menu 
sections above, before proceeding. 
Allow the analyzer reading to stabilize, Advance the cursor (*) to the “Span” option as illustrated and press the green ENTER 
key to accept the selection. 
 
MAIN MENU
 Sample 
* Span 
 Zero 
 Alarm 
 System 
 Standby 
 
Sample
1.00 PPM
Auto Range 
85⁰F 100Kpa 
0 to 10 PPM 
12/31/07 12:00:00 
 
The following menu appears: 
 
SPAN Sample
* Factory Default 
 Calibrate 
 
 
 
 
 
 
Auto Range 0 to 10 PPM 
85⁰F 100Kpa 12/31/07 12:00:00 
 
Advance the cursor (*) to the Factory Default Span option and press ENTER. 
Within seconds the system returns to the MAIN MENU in the “Sample” mode. 
 
MAIN MENU
 Sample 
* Span 
 Zero 
 Alarm 
 System 
 Standby 
 
Sample
1.00 PPM
Auto Range 
85⁰F 100Kpa 
0 to 10 PPM 
12/31/07 12:00:00 
 
At this point, the user must decide whether to perform a Span Gas or Air Calibration, see above. 
 
Note: If the analyzer is to calibrated with instrument or compressed air follow the Span Gas Calibration procedure below. 